Guide Camera Removal
Last modification: Sept. 10, 2007
The guide camera cannot be removed or installed while the spectrograph is mounted on the telescope.The guide camera does not have to be removed from the instrument for shipping.
EQUIPMENT
SHORTING PLUG
Grounding Strip
PROCEEDURE
1. The guide camera does not need to
be removed from the instrument for shipping.
2. The electronics connections do need
to be removed for shipment.
3. Plug in the grounding strip to
outlet.
4. put on grounding bracelet and push
up arm until full connection is made.
5. Turn off power to the black
CCD electronics box.
6. Disconnect the small circular
connector on the back of the guide camera.
7. Disconnect the hoses from the back
of the guide camera. These remove by pushing in the metal tabs for
them to pop out.
8. Get the shorting plug.
9. Disconnect the main electronics
cable and replace with shorting plug on the back of the camera.
10. Roll back cables into the box and
coil neatly.
11. Coil balance of cables (main AC,
Ethernet, etc. into the black electronics box.)